Example of a Standard Curve for Molecular Mass
Protein standards for gels are purified polypeptides
with relative mobilities that correspond closely
to their true molecular mass. Suppliers of chemicals
for electrophoresis such as SIGMA (St. Louis, MO)
or Bio-Rad (Hercules, CA) provide ready-made molecular
mass standards. SIGMA sells standards for calibrating
SDS gels with a tris based buffer system (Laemmli
gels).
Inserts with commercial standards, older literature,
and even my old web pages may refer to molecular
weight (MW) rather than molecular mass. MW is the
same as relative molecular mass (Mr) but differs
from molecular mass in that it is a unitless quantity.
You will likely find MW data
that are reported in Daltons or kiloDaltons although such usage
is not correct. A number representing MW is identical
to the corresponding molecular mass in Daltons,
thus the terms can be used interchangeably as long
as the units are dropped from quantities that are
represented as MW.
SIGMA Standard Mixture for Molecular Weights 30,000-200,000
(SDS6H2)
- myosin from porcine muscle, 200,000
- beta-galactosidase, from Escherichia
coli, 116,000
- phosphorylase B, from rabbit muscle 97,400
- albumin, bovine 66,000
- albumin, from chicken egg white 45,000
- carbonic anhydrase from bovine erythrocytes,
29,000
SIGMA Dalton Mark VII-L Standard Mixture, MW range
14,000-70,000 (SDS-7)
- albumin, bovine 66,000
- albumin, from chicken egg white 45,000
- glyceraldehyde-3-phosphate dehydrogenase, from
rabbit muscle 36,000
- carbonic anhydrase from bovine erythrocytes,
29,000
- trypsinogen, from bovine pancreas 24,000
- trypsin inhibitor, soybean, 20,100
- alpha-lactalbumin, bovine milk 14,200
Note that the sources of proteins are varied.
You won't find all of them in any one protein fraction,
in fact you aren't likely to find any of them,
depending on the fraction you are studying. They
are used to calibrate gels, not as indicators of
what types of proteins are present. Keep in mind
that many very different proteins have similar
molecular weights. The patterns given by standard
mixes become recognizable with experience.
A typical plot of the molecular weights of standards
versus their relative mobilities is shown below,
using a log scale for the molecular weights. Be
very careful with curve fits, keeping in mind that
the scale is logarithmic. You don't want to be
in error near the top of the gel. It may be better
to simply interpolate results (connect the data
points).
Relative mobility for a given polypeptide will
vary with gel density (%T).
Curves will always shift when percent acrylamide
changes.
With some gels of high percent acrylamide,
the dye front will not be evident, since proteins
small enough to run with the same mobility as the
bromphenol blue dye may not be present. Unless
the position of the dye front was marked before
staining, a true relative mobility cannot be determined.
As long as a gel is calibrated using internal standards,
MW estimates can be obtained by representing relative
migration distance using an arbitrary reference
point such as the bottom of the gel. Use of a true
relative mobility allows one to use the same standard
curve for any gel of the exact same composition
regardless of dimensions or position of the dye
front when electrophoresis is terminated. Error
in MW estimates is compounded with the use of an
arbitrary reference point when the original dye
front was curved or otherwise distorted.
|