Principles of Spectrophotometry
A spectrophotometer consists of two instruments,
namely a spectrometer for producing
light of any selected color (wavelength), and
a photometer for measuring the intensity
of light. The instruments are arranged so that
liquid in a cuvette can be placed between the
spectrometer beam and the photometer. The amount
of light passing through the tube is measured
by the photometer. The photometer delivers a
voltage signal to a display device, normally
a galvanometer. The signal changes as the amount
of light absorbed by the liquid changes.
If development of color is linked to the concentration
of a substance in solution then that concentration
can be measured by determining the extent of
absorption of light at the appropriate wavelength.
For example hemoglobin appears red because the
hemoglobin absorbs blue and green light rays
much more effectively than red. The degree of
absorbance of blue or green light is proportional
to the concentration of hemoglobin.
When monochromatic light (light of a specific
wavelength) passes through a solution there is
usually a quantitative relationship (Beer's law)
between the solute concentration and the intensity
of the transmitted light, that is,
where I sub 0 is the intensity of transmitted
light using the pure solvent, I is the intensity
of the transmitted light when the colored compound
is added, c is concentration of the colored compound,
l is the distance the light passes through the
solution, and k is a constant. If the light path
l is a constant, as is the case with a spectrophotometer,
Beer's law may be written,
where k is a new constant and T is the transmittance
of the solution. There is a logarithmic relationship
between transmittance and the concentration of
the colored compound. Thus,
The O.D. is directly proportional to the concentration
of the colored compound. Most spectrophotometers
have a scale that reads both in O.D. (absorbance)
units, which is a logarithmic scale, and in %
transmittance, which is an arithmetic scale.
As suggested by the above relationships, the
absorbance scale is the most useful for colorimetric
assays.
Using a Spectronic 20 spectrophotometer
The Spectronic 20 spectrometer is widely used
in teaching laboratories. The specific instructions
will differ with other models, but the principles
remain.
- The instrument must have been warm for at
least 15 min. prior to use. The power switch
doubles as the zeroing control.
- Use the wavelength knob to set the desired
wavelength. Extreme wavelengths, in the ultraviolet
or infrared ranges, require special filters,
light sources, and/or sample holders (cuvettes).
- With the sample cover closed, use the zero
control to adjust the meter needle to "0" on
the % transmittance scale (with no sample in
the instrument the light path is blocked, so
the photometer reads no light at all).
- Wipe the tube containing the reference solution
with a lab wipe and place it into the sample
holder. Close the cover and use the light control
knob to set the meter needle to "0" on the
absorbance scale.
- Remove the reference tube, wipe off the first
sample or standard tube, insert it and close
the cover. Read and record the absorbance,
not the transmittance.
- Remove the sample tube, readjust the zero,
and recalibrate if necessary before checking
the next sample.
Why use a reference solution? Can't you just
use a water blank? A proper reference solution
contains color reagent plus sample buffer. The
difference between the reference and a sample
is that the concentration of the assayable substance
in the reference solution is zero. The reference
tube transmits as much light as is possible with
the assay solution you are using. A sample tube
with any concentration of the assayable substance
absorbs more light than the reference, transmitting
less light to the photometer. In order to obtain
the best readability and accuracy, the scale
is set to read zero absorbance (100% transmission)
with the reference in place. Now you can use
the full scale of the spectrophotometer. If you
use a water blank as a reference, you might find
that the assay solution alone absorbs so much
light relative to distilled water that the usable
scale is compressed, and the accuracy is very
poor.
|