GPC Operation
1. Launch
Breeze software.
·
If Windows background screen is showing, click on
“Breeze” icon. Close “Waters Breeze
Help” display.
2. Look
at the Flow meter. Confirm that the flow
is 0.1 mL/min. If it is not, find Dr.
Saterbak.
3. Look
at the Pressure meter. Confirm that the pressure is <1000 psi. If it is not, find Dr. Saterbak.
4. Locate
carousel. It may be on the counter or in
the chamber of the Autosampler labeled “Carousel”.
5. Load
1 mL vials into the carousel beginning at position 1 and continuing to the
left.
6. Place
carousel in Autosampler.
7. Click
on “Manage Breeze” button.
8. Click
on “Change Project/User” box.
9. Select
your group name from the list of projects.
Click OK. The screen that appears
next will have the “Find Data” button depressed and a set of tabs available.
10. Click
on “Sample Queue” button.
11. On
pull-down menu, go to File à New Sample Set à Using Sample
Set Method.
12. To
run samples, select “BIOE 441 Samples”.
Click “Open”.
13. The
screen that appears next will contain the following chart. Columns further to the right of Injection
Volume exist but should not be changed.
|
Vial |
Sample Name |
Function |
Method |
Run Time
(min) |
Injection
Volume (mL) |
1 |
|
|
Clear Calibration |
PLLA/PLGA method |
|
|
2 |
|
|
Condition Column |
Ramp from 0_1 |
30 |
|
3 |
|
|
Equilibration |
Equilibrate at 0_75 |
60 |
|
4 |
1 |
X |
Inject Broad Samples |
PLLA/PLGA method |
20 |
100 |
5 |
2 |
X |
Inject Broad Samples |
PLLA/PLGA method |
20 |
100 |
6… |
3 … |
X |
Inject Broad Samples |
PLLA/PLGA method |
20 |
100 |
Y |
Z |
X |
Inject Broad Samples |
PLLA/PLGA method |
20 |
100 |
|
|
|
|
|
|
|
where X is the Sample Name that
must be entered,
Y is the number of
command lines, and
Z is the number of total
vials that are being tested.
14. Enter
your sample names, X.
15. Confirm
that the number of vials in the chart EXACTLY corresponds to the number of
vials you are testing. Injecting air
into the column by running a “non-existent” sample can cause expensive damage
to the column.
·
If there are too many sample vials, delete a row
(from the bottom) by right clicking in the Function column and selecting the
Delete Rows option.
·
If there are not enough rows, right click in the Function column on the
first empty line. A line duplicating the
one above should be added. Make sure
that the information in the columns on Vial Number, Sample Name, Function,
Method, Run Time and Injection Volume are correct.
16. Check
with Dr. Saterbak to confirm your sample protocol.
17. On
the pull-down menu, go to File à Save Sample Set Method to save
the updated method set.
18. On
the pull-down menu, go to Inject à Run Sample Set.
·
Type in a name for your sample set such as “BIOE
441 9_23_01”.
·
For the settings, select “Run and Process”.
·
Hit “Run”.
19. The
“Current Sample Set Method” icon (3 test tubes and green triangle) should go
gray. The “Abort Run” icon (3 test tubes
and red square) should be available.
Typically, your samples will take 2-5 hrs to run. They can run without monitoring during the lab
session and overnight. After the last
sample, the pump will automatically ramp down to 0.1 mL/min.
When your samples are running, the data is graphically reported in a
plot of mV versus minutes at the bottom of the computer screen.
On Monday of each week, Dr. Saterbak will run a set of polystyrene
sample. She will post the retention
times for compounds with the following molecular weights (g/mol):
4,000 20,000 50,000 100,000 1,000,000
Use this data to develop a mathematical relationship between retention
time and molecular weight.
Method Parameter Settings
·
Run time of 20 min for samples.
·
Flow rate of 0.75 mL/min for samples.
·
Sample injection volume of 100 mL.
·
Gradient flow.
·
High-pressure limit of 1000 psi.
·
RI detector with sampling rate of 5/sec, filter
time of 3, + polarity, mV units, sensitivity of 64.
·
Internal and external 1 temperatures set at 30 °C.
The GPC runs at 0.1 mL/min during times when samples are NOT running
(e.g. overnight).
GPC Data Analysis
1. Launch
Breeze software.
·
If Windows background screen is showing, click on
“Breeze” icon. Close “Waters Breeze
Help” display.
2. Click
on “Find Data” button. Click on
“Channels” tab.
·
If you are running samples and want to look at the
results so far, click on the “Update” button.
·
All of the samples that you have run should be
listed here, sorted by Date Acquired.
3. Select
a run to analyze by clicking once on the row of interest. To select multiple rows, hold the control
button down and click on the rows of interest.
4. With
cursor on row(s) of interest, right click and select “Review.”
·
Selected rows should appear as list under “View
Data” button.
·
Additional rows may be added to the “View Data”
list by repeating Steps 2.-3.
5. The
GPC chromatogram for a particular sample can be seen by clicking on the number
in the first column of the table.
6. To
determine the Retention Time, hit the Integrate icon (graph with two peaks and
a liquid drop) or go to the pull-down menu and select Process à Integrate.
·
The Retention Time (in minutes) should appear on
the graph. Record the Retention Time.
·
If the integration ends (marked with red triangles)
are not in a technically sound location, contact Dr. Saterbak. The integration ends can be adjusted.
·
The Retention Times should be between 8 min and 14
min. Water and other small molecules
exit the column around 15-18 min. Peaks
above and below the baseline in this area should be ignored.
·
If you want to clear the integration and repeat, go
to the pull-down menu and select Edit à Clear Integration. You can then integrate the refreshed curve.
7. Repeat
Steps 5. - 6. for each run of interest.
8. To
look at all your chromatograms at once, click on the Overlay icon (graph with 2
traces, each with 2 peaks) or go to the pull-down menu and select Plot à Overlay.
9. Once
data analysis is completed, go to the pull-down menu and select File à Save à Result. If you ever see a Breeze box asking “Do you
wish to save before switching,” answer “Yes.”
Occasionally you will have difficulty saving because of a partially
collected chromatogram or other issues; just follow through with “OK”.